CN

EN108 Fully Automated Library Preparation System

EN108 Fully Automated Library Preparation System
INTRODUCTION
EN108, is a fully automatic pipetting workstation integrated pipetting module, magnetic purification module, shaker module, PCR module, thermal incubation module and refrigeration module, which can be used for high-throughput sequencing library preparation of DNA and RNA samples.
Instrument characteristics
Highly Automated Highly Automated
Highly Automated
Truly walk-away
Built-in code scanning and data recording system to record and track data information.
High efficiency High efficiency
High efficiency
It operates uninterrupted for 24 hours and processes up to 32 samples per run.
High Safety Factor High Safety Factor
High Safety Factor
Prepare liquid drip-proof pipetting module, standard filter cartridge tip, built-in high-efficiency filtration system and sterile ultraviolet lamp, and ensure that the whole experimental process is clean and pollution-free from multiple angles.
High Accuracy High Accuracy
High Accuracy
Automated standard operating procedures effectively eliminate manual errors and improve the stability and repeatability of results.
High Flexibility High Flexibility
High Flexibility
Adaptable library preparation kits for multiple brands and multiple sequencing types.
Customized service can be completed according to user needs.

Modules

Product parameters

Size and Weight

Dimensions

900mm(W)*810mm(D)*900mm(H)

Net weight

172.5kg

Power socket

AC Voltag

220V

AC frequenc

50Hz

Max Power

1300VA

Pipetting performance

Pipetting principle

Air displacement

Pipetting rang

2μl-200μl

Pipetting homogeneity (cv)

2μl:<5%
200μl:<1%

Pipetting accuracy

2µl:<±5%
200μl:<±1%

Pipette Arm Positioning Accuracy

±0.1mm

PCR Instrument

Temperature range

4~99℃

Temperature homogeneity

≤±0.5℃

Temperature accuracy

≤±0.3℃

Refrigeration Module

Temperature range

2-7℃

Temperature accuracy

±1℃ at 4℃

Temperature homogeneity

±1℃ at 4℃

Thermal incubation module

Temperature range

Room temperature ~ 70℃

Temperature accuracy

±1℃ at 65℃

Temperature homogeneity

±1℃ at 65℃

Safety protection

Positive Pressure HEPA SystemFilter Efficiency ≥99.99% at 0.3μm, with Work Area  Cleanliness Level Achieving ISO Grade 5

UV lamp

Radiation Dose in 20 minutes Exceeds 100,000 µW·s/cm²


Related images

Make it easier for more individuals and organizations

Copyright © Established Nayo Biotechnology (Shanghai) Co., Ltd. All Rights Reserved.

沪ICP备2021024954号 XML Web design: Surfer Technology